Second Quarter 2005
INL Quarterly Site Environmental Report
Return to Index
The ESER Quality Assurance Program consists of five ongoing tasks which measure:
- method uncertainty;
- data completeness;
- data accuracy, using spike and laboratory control samples;
- data precision, using split samples, duplicate samples, and recounts; and
- the presence of contamination in samples, using blanks.
The following discussion summarizes the results of the quality assurance program for the period from April 1 to June 30, 2005.
The Quality Assurance Project Plan (QAPP) establishes data quality and method quality objectives for the ESER surveillance program (Stoller 2002). Since the primary concern is with detection, the lower bound for the method uncertainty is set at zero. The upper bound is defined by the ESER program as the maximum concentration in the non-outlier range of data from the past seven years. Each individual result is checked for acceptance on the basis of the result, whether it is below the lower limit (i.e., a negative value), greater than the upper limit, or between the lower and upper limit (the most common occurrence). The calculated method uncertainty is then compared to the 1s measured uncertainty. A sample is deemed acceptable when the measured 1s uncertainty is less than the calculated uncertainty. The upper bound values are currently being evaluated and revised. Preliminary results indicate that more calculated method uncertainties for detected results were acceptable.
The Quality Assurance Project Plan (QAPP) specifies a 98 percent completeness goal for all regularly scheduled sample types (Stoller 2004). Data completeness for sample collection and delivery was 100 percent during the second quarter for all sample types with these exceptions: a number of precipitation samples were not collected due to lack of precipitation and one (2 percent) of the scheduled PM10 samples did not run when the timer did not trip.
Five air samples were determined to invalid due to insufficient volume collected because of equipment failure or electrical work (one from Arco and two each from Dubois and Howe). The completeness of air filter data is thus considered to be 98.1percent.
Accuracy is a measure of the degree to which a measured value agrees with the "true" value for a given parameter; accuracy includes elements of both bias and precision.
During the second quarter of 2005, spikes (samples prepared with known amounts of radionuclides) of the following types were obtained and submitted:
A total of 16 values for comparison were generated by the analysis of the spike samples. All results were within the criteria during the second quarter except for Cobalt-60 in milk and Barium-133 in the charcoal cartridge.
The Quality Assessment Program was discontinued following the March 2004 distribution. Performance evaluation samples are now prepared through the Mixed Analyte Performance Evaluation Program (MAPEP), administered by the Department of Energy’s Radiological and Environmental Sciences Laboratory. DOE has mandated that all laboratories performing analyses in support of the Office of Environmental Management shall participate in MAPEP. The program distributes samples of air, water, vegetation and soil for analysis in approximately January and June. Both radiological and nonradiological constituents are included in the program.
Both the Idaho State University EAL and Severn-Trent participated in the program in May 2005. Results are tabulated below for those analyses performed by each laboratory for the ESER program. (A = Acceptable, W = Acceptable with warning, N = Not acceptable)
Idaho State University Environmental Assessment Laboratory |
||||||
|
Matrix: Air (Bq) |
||||||
|
Analyte |
EAL Result |
MAPEP Result |
Bias |
Acceptable Range |
Evaluation |
|
|
Cesium-134 |
2.9 |
3.51 |
-17.4 |
2.46-4.56 |
A |
|
|
Cesium-137 |
2.0 |
2.26 |
-11.5 |
1.58-2.94 |
A |
|
|
Cobalt-57 |
4.5 |
4.92 |
-8.5 |
3.44-6.40 |
A |
|
|
Cobalt-60 |
2.8 |
3.03 |
-7.6 |
2.12-3.94 |
A |
|
|
Manganese-54 |
3.0 |
3.33 |
-9.9 |
2.33-4.33 |
A |
|
|
Zinc-65 |
2.9 |
3.14 |
-7.6 |
2.20-4.08 |
A |
|
|
Gross alpha |
0.09 |
0.232 |
-61.2 |
0.000-0.464 |
A |
|
|
Gross beta |
0.32 |
0.297 |
7.7 |
0.148-0.446 |
A |
|
|
Matrix: Water (Bq/L) |
||||||
|
Cesium-134 |
108.9 |
127 |
-14.3 |
88.9-165.1 |
A |
|
|
Cesium-137 |
317.2 |
332 |
-4.5 |
232.4-431.6 |
A |
|
|
Cobalt-57 |
223.3 |
227 |
-1.6 |
158.9-295.1 |
A |
|
|
Cobalt-60 |
252.4 |
251 |
0.6 |
175.7-326.3 |
A |
|
|
Manganese-54 |
326.9 |
331 |
-1.2 |
231.7-430.3 |
A |
|
|
Zinc-65 |
532.1 |
496 |
7.3 |
347.2-644.8 |
A |
|
|
Gross alpha |
0.11 |
0.525 |
-79.0 |
0.000-1.050 |
A |
|
|
Gross beta |
2.00 |
1.67 |
19.8 |
0.835-2.505 |
A |
|
|
Matrix: Soil (Bq/kg) |
||||||
|
Cesium-134 |
609.6 |
759 |
-19.7 |
531.3-986.7 |
A |
|
|
Cesium-137 |
265.8 |
315 |
-15.6 |
220.5-409.5 |
A |
|
|
Cobalt-57 |
206.9 |
242 |
-14.5 |
169.4-314.6 |
A |
|
|
Cobalt-60 |
190.7 |
212 |
-10.0 |
148.4-275.6 |
A |
|
|
Manganese-54 |
442.0 |
485 |
-8.9 |
339.5-630.5 |
A |
|
|
Potassium-40 |
515.6 |
604 |
-14.6 |
422.8-785.2 |
A |
|
|
Zinc-65 |
737.7 |
810 |
-8.9 |
567.0-1053.0 |
A |
|
Severn-Trent Laboratory |
||||||||||
|
Matrix: Air (Bq) |
||||||||||
|
Analyte |
S/T Result |
MAPEP Result |
Bias |
Acceptable Range |
Evaluation |
|||||
|
Americium-241 |
0.109 |
0.102 |
6.9 |
0.07-0.13 |
A |
|||||
|
Plutonium-238 |
0.201 |
0.195 |
3.1 |
0.14-0.25 |
A |
|||||
|
Plutonium-239/240 |
0.17 |
0.165 |
3.0 |
0.12-0.21 |
A |
|||||
|
Strontium-90 |
1.36 |
1.35 |
0.7 |
0.94-1.75 |
A |
|||||
|
Matrix: Water (Bq/L) |
||||||||||
|
Americium-241 |
1.60 |
1.72 |
-7.0 |
1.20-2.24 |
A |
|||||
|
Plutonium-238 |
0.024 |
0.018 |
a |
|
A |
|||||
|
Plutonium-239/240 |
2.50 |
2.4 |
4.2 |
1.68-3.12 |
A |
|||||
|
Strontium-90 |
0.0b |
|
|
|
W |
|||||
|
Matrix: Soil (Bq/kg) |
||||||||||
|
Americium-241 |
106 |
109 |
-2.8 |
76.3-141.7 |
A |
|||||
|
Plutonium-238 |
0.67 |
0.48 |
a |
|
A |
|||||
|
Plutonium-239/240 |
90.2 |
89.5 |
0.8 |
62.65-116.35 |
A |
|||||
|
Strontium-90 |
631 |
640 |
-1.4 |
448-832 |
A |
|||||
|
Matrix: Vegetation (Bq) |
||||||||||
|
Americium-241 |
0.234 |
0.145 |
61.4 |
0.10-0.19 |
N |
|||||
|
Plutonium-238 |
0.36 |
0.224 |
60.7 |
0.16-0.29 |
N |
|||||
|
Plutonium-239/240 |
0.0b |
0.0006 |
|
|
W |
|||||
|
Strontium-90 |
2.1 |
1.65 |
27.3 |
1.15-2.14 |
W |
|||||
|
a. Result was a statistical nondetect. |
||||||||||
|
b. Reported zero uncertainty. |
||||||||||
The Idaho State University Environmental Assessment Laboratory uses NIST standards to prepare spiked water samples and uses commercially prepared calibration standards as NIST-traceable spiked samples. ISU considers a performance to be acceptable if results pass either the ±20 percent test specified by the ESER program or the three-sigma test described in the data precision section. A variety of checks are made each quarter on different geometries.
During the second quarter of 2005, 19 analyses were conducted on NIST-traceable
standards for gamma-emitting radionuclides. Geometries tested included
low-volume air filter composites, single charcoal cartridge screening,
10-charcoal cartridge screening, 500 ml 1.0 g/cc samples, and one-liter 1.0 g/cc
samples. A total of 115 analytical results were generated. All of the results
were within the ±20 percent range.
Water samples spiked with tritium received 14 analyses during the quarterly
reporting period. All were well within the ±20 percent criterion, and in fact
all were within 7 percent of the known value. Gross beta spikes analyzed in the
first quarter were within 20 percent of the expected values; two of four gross
alpha spikes were within 20 percent and all were within three standard
deviations. An internal tritium milk spike was within approximately 2 percent of
the known value.
Severn-Trent analyzes a laboratory control sample (LCS) with each batch of samples submitted by the ESER. During the second quarter this consisted of strontium-90 and actinides in air and strontium-90 in milk.
|
Media |
Analyte |
QAPjP Accuracy |
LCS Result |
Within Criterion? |
|
Air |
Strontium-90 |
±10 percent |
+13.3 percent |
No |
|
Air |
Americium-241 |
±10 percent |
-12.4 percent |
No |
|
Air |
Plutonium-239/240 |
±10 percent |
+3.9 percent |
Yes |
|
Milk |
Strontium-90 |
±25 percent |
+2.2 percent |
Yes |
Data precision is a measure of the variability associated with a measurement
system. Precision is measured using duplicate samples, split samples, and
recounts. Data precision is measured using duplicate samples, split samples, and
recounts. The Quality Assurance Project Plan specifies that sample results
should agree within ±20 percent or 3σ, whichever is greater. For environmental
samples at levels that are within the normal range found by the ESER, the 3
standard deviation criterion is the one that applies in nearly all cases. The
standard deviation criterion is considered to be met if the values of the
duplicate samples differ by less than the root mean square of three standard
deviations of each sample result. Mathematically, this is expressed as:
│X-Y│< 3 (sqrt(σx2 + σy2)),
where:
X is the result of the regular sample
Y is the result of the duplicate sample
σx is the uncertainty of the regular sample
σy is the uncertainty of the duplicate sample
Another measure of duplicate sample results is the relative percent difference.
This value is the difference in the two results divided by the mean of the two
results. The following sections of this report first check the sample results
using the 3 standard deviation criterion. If this criterion is not met, the
results are then listed for the relative percent difference.
Duplicate milk samples were collected from Moreland on April 5, Dietrich on
June 7, and Howe on June 7 and analyzed for gamma-emitting radionuclides. All
results were within the 3σ criteria.
Duplicate water samples were collected from Arco (drinking water) and Bliss
(surface water) and analyzed for gross alpha, gross beta, and tritium. All
results were within the 3σ criteria.
Duplicate air samplers are operated at two locations adjacent to regular air samplers. In the first quarter of 2005 these samplers, designated as QA-1 and QA-2, were in operation at the INL Main Gate and Howe, respectively. Particulate filters receive the standard analysis for gross alpha and gross beta; charcoal cartridges are analyzed specifically for iodine-131. All gross alpha and gross beta results for the co-located samplers met the acceptability criteria. Charcoal cartridge results are difficult to present because cartridges are counted in batches of ten.
Composite air samples from the two QA samplers were submitted for analysis at the end of the second quarter for gamma spectrometry at the EAL and for 90Sr at Severn-Trent. All analyses were within the 3s criterion.
A comparison of duplicate results can also show bias in the sampling system.
For example, if one set of results is consistently lower or higher than the
other one might suspect that this bias was due to a leak in the system or
variations in the calibration of the flow meter. Figure 15
and Figure 16. show the
difference in results (Main sampler - QA duplicate sampler) over time. The
figures show that the bias is small and not consistent in one direction,
indicating that there is no obvious bias in the duplicate sampling systems in
these cases.
The EAL splits and analyzes a number of milk, precipitation, and atmospheric moisture samples each quarter. The laboratory tests each result using both the ±20 percent criterion and the 3s criterion, although it considers the former test meaningless for analyses producing fewer than 15 total counts and questionable even where counts are on the order of 100. The latter criterion is applied in nearly all cases at the levels seen in environmental samples analyzed for the ESER program. Results of the EAL split sample analyses met the criteria for acceptance during the second quarter 2005.
The ISU EAL recounts a number of samples of each media type. The lab tests each recount using both the 20 percent criterion and the 3σ criterion, subject to the limitations described in the previous section.
A summary of the recount results for the second quarter is presented below.
The ESER program submits field blanks along with the regular samples to test for the introduction of contamination during the process of field collection, laboratory preparation, and laboratory analysis. The current program includes the use of two field blanks, designated as Blank A and Blank B, that each accompanies one of the air filter routes. Quarterly composites of the blanks are also submitted. After gamma spectrometry analysis, one of the blanks is analyzed for Sr-90 and the other for transuranics.
The Quality Assurance Project Plan does not specify requirements for blank
performance, but ideally the result should be within ±2σ of zero and preferably
within ±1σ of zero on most analyses. It would be expected, based on counting
statistics for a sample that was truly a blank (i.e., the true value of the
analyte was zero), that 68.3 percent of analyses would fall within one standard
deviation, 95.5 percent would fall within two standard deviations, and 99.7
percent would fall within three standard deviations. With a single exception in
a gross beta analysis, all results were within the 3σ significance level.
The Environmental Assessment Laboratory prepares and analyzes reagent blanks to help determine if the analysis will yield a zero result when no activity is present. ISU considers the result within specification if the concentration is less than the minimum detectable concentration (MDC) for the analysis. One such blank was analyzed for tritium in the second quarter for milk. The blank was below the MDC for the analysis and less than two standard deviations. A water blank analyzed for gross alpha and gross beta was also below the MDA for the analysis and within two standard deviations for both parameters.
Severn-Trent analyzes a blank with each set of results. Second quarter blanks were less than three standard deviations of zero for strontium-90, plutonium-238, plutonium-239/240 and americium-241 in air and for strontium-90 in milk.