|
If
the INEEL was a significant source of offsite contamination,
concentrations of contaminants would be higher at INEEL locations
compared to Boundary locations which, in turn, would be higher than
Distant locations. An
independent samples t-test (a=
0.05) was used to determine if there were statistically significant
differences between the average gross alpha and gross beta
concentrations at INEEL, Boundary, and Distant locations INEEL, Boundary, and Distant location weekly average gross alpha concentrations in air are shown in Figure 8. Weekly average gross beta concentrations are shown in Figure 9. During two separate weeks (ending April 11, and May 23), the average gross beta concentration for INEEL locations was significantly higher than at Distant locations. Also during two separate weeks (ending April 18 and May 23), the average gross beta concentration for INEEL locations was significantly higher than at Boundary locations. During the week of April 11, the average gross beta concentration was higher at Boundary locations than at Distant locations. During
two separate weeks, those ending April 4, and June 20, the average gross
alpha concentration for INEEL locations was significantly higher than at
Distant locations. During
those same weeks, the average gross alpha concentration at Boundary
locations was significantly higher than Distant locations.
There were no consistent trends over time of INEEL locations being higher than Boundary locations, being higher than Distant locations as one would expect if the INEEL was the source of radionuclide contamination. Gross alpha and gross beta results for individual filters are listed in Table C-1 of Appendix C. Monthly average gross alpha and beta concentrations in air at each sampling location are shown in Figures 10 – 15.
Of
the nine samples submitted for analysis, 241Am was detected
in five of them. Four of
the five samples had levels greater than their respective two standard
deviation (2s) and MDC (see Table 2).
One sample, from the Arco (QA-1) station, had an 241Am
result greater than 2s, but less than the MDC, indicating it is most
likely a false positive. 239/240Pu
was detected in four of the samples submitted.
Three of the four samples had levels greater than their
respective 2s and MDCs (see Table 2).
One sample, from the Main Gate location, had a result greater
than 2s, but less than the MDC, indicating it is most likely a false
positive. TABLE 2. Specific radionuclides with results > 2s and results > MDC in composite air filters.
90Sr
was detected in two of the samples submitted at a level greater than
their associated 2s values. However,
both were less than the MDCs, indicating they are most likely false
positives. Since 239/240Pu and 241Am were deposited around the world from atmospheric nuclear weapons testing the detected amounts do not necessarily indicate inputs from the INEEL, though this cannot be ruled out. However, concentrations are within the range of values measured throughout the World (EPA, 2002). In addition, detected values in composite samples for 241Am ranged from 5,100 to over 10,500 times smaller than the DCG value of 2 x 10-14 μCi/mL. Values detected in composite samples for 239/240Pu ranged from 5,800 to over 17,500 times smaller than the DCG value of 2 x 10‑14 μCi/mL. Results for composite filter samples from the second quarter, 2001, are shown in Table C-3 of Appendix C. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||